2
dc volts
5 ranges: 0.1 V to 1000 V
8.5 to 4.5 digit resolution
Up to 100,000 readings/sec
(4.5 digits)
Maximum sensitivity: 10 nV
0.6 ppm 24 hour accuracy
8 ppm (4 ppm optional)/year
voltage reference stability
Ohms
9 ranges: 10 Ω to 1 GΩ
Two-wire and four-wire ohms with
offset compensation
Up to 50,000 readings/sec
(5.5 digits)
Maximum sensitivity: 10 µΩ
2.2 ppm 24 hour accuracy
ac volts
6 ranges: 10 mV to 1000 V
1 Hz to 10 MHz bandwidth
Up to 50 readings/sec with all
readings to specified accuracy
Choice of sampling or analog true
rms techniques
100 ppm best accuracy
dc current
8 ranges: 100 nA to 1 A
Up to 1,350 readings/sec (5.5 dig-
its)
Maximum sensitivity: 1 pA
14 ppm 24 hour accuracy
ac current
5 ranges: 100 µA to 1 A
10 Hz to 100 kHz bandwidth
Up to 50 readings/sec
500 ppm 24 hour accuracy
Frequency and period
Voltage or current ranges
Frequency: 1 Hz to 10 MHz
Period: 100 ns to 1 sec
• 0.01% accuracy
ac or dc coupled
Maximum speeds
100,000 readings/sec at 4.5 digits
(16 bits)
50,000 readings/sec at 5.5 digits
6,000 readings/sec at 6.5 digits
60 readings/sec at 7.5 digits
6 readings/sec at 8.5 digits
Measurement set-up speed
100,000 readings/sec over GPIB or
with internal memory
• 110 autoranges/sec
340 function or range changes/sec
Post-processed math from internal
memory
Performance Highlights

Other manuals for Agilent Technologies 3458A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 3458A and is the answer not in the manual?

Agilent Technologies 3458A Specifications

General IconGeneral
BrandAgilent Technologies
Model3458A
CategoryMultimeter
LanguageEnglish

Related product manuals