Mountingandcommissioning
©byBlum‐Novotest
17
V2A|04/2018
1. Insertthemeasuringsystemintothetoolholder(systemBTH).
2. Tightentheclampingscrewjustslightly.
3. Inserttheprobeintothespindle(TC50,TC52,TC53‐10,TC53‐30,TC54‐10only).
4. Setthedialgaugeonthestylus.
5. Rotatetheprobeby360°.
6. Adjusttheconcentricity(4hexagonsocketscrew s2mm):
e.g.X‐dir
ection:
Releasethecentringscrews(2a)and(2c)justslightly.
Adjusttheconcentricitybyalternatereleaseandtighteningofthecentringscrews(2a)and
(2c).
Y‐direction:
Releasethecentringscrews(2b)and(2d)justslightly.
Adjusttheconcentricitybyalternatereleaseandtighteningofthecentringscrews(2a)and
(2c).
Concentricitydeviation:
TC50,TC52,TC
53‐10,TC
53‐30,TC54‐10:<10µm
TC51:<100µm
Allcentringscrewsaretightened.
7. Tightentheclampingscrew.
Theconcentricityadjustmentisnowfixed.
8. Checktheconcentricity.
9. Carryoutacalibrationcycle.
Adjustmentoftherun‐outatTC51withoptic alpre set ter!