10093955
1/10
Package Should Contain:
1. Proportioner unit.
2. Supply tubing 28' total.
3. Foot valve assemblies &
weights (4).
4. Discharge tubing for
each eductor.
5. Metering tip kits.
6. Mounting anchor kit.
7. Hose hook for 3.5 GPM
eductor.
8. Instruction sheet.
Clip or
bend
these
tabs to
depress button
into locked
position.
Proportioner
Model 39741 & 39751
with E-gap Eductors
Hydro Systems 3798 Round Bottom Road, Cincinnati, OH 45244 ▲ Phone: (513) 271-8800 ▲ Fax: (513) 271-0160
Installation and Operation:
1. Find suitable place close to water source for unit. Mounting bracket should be installed approx. 5' from oor. Level
bracket & mark holes. Drill 9/32" holes & install mounting anchors and screws in bracket. Drill holes for the three wall
anchors with a 5/16" drill bit, using the cabinet back as template for proper spacing of the mounting screws. Install
mounting anchors, and then screws in top two anchors. Slide key holes in cabinet back over screw heads, tighten
screws, then install bottom screw. Do not mount more than 6 feet (1.8 meters) above the bottom of the concentrate
container, nor below the highest concentrate level (never mount your concentrate higher than the proportioner).
2. Select metering tips (up to 4) for each selector valve. (see next two sections) Push each tip rmly into a separate hose
barb extending from the selector valve. A tip with no hole (clear plastic) can be used to block any valve port being
used. (This may be used for dispensing water only). Select and install a metering tip for the single product eductor
(right side) in the same manner.
3. Cut tubing provided into separate supply tubes for each product to be dispensed (tubing allows 7 feet of tube per
product). Supply tubes should reach from hose barbs on the selector valve body and eductor to bottom of concentrate
containers. Slide ceramic weights over one end of each tube and slide foot valves into the same ends of the tubes.
4. Slip other end of supply tube through an opening in either side of the cabinet and push over the hose barb/metering
tip on the eductor. (Repeat for all eductors.)
5. Place foot valve ends of supply tubes into concentrate containers. REMEMBER TO CHECK FOOT VALVE STRAINERS
PERIODICALLY FOR CLOGGING: CLEAN IF NECESSARY.
6. A short discharge tube is used with the 1GPM eductor; minimum tube length is 8 inches (20cm) for proper opera-
tion. Longer tubes (4 feet) are used with a 3.5 GPM eductor. Do not remove the ooding rings from inside the tubes.
Slide end of tube with ooding ring over eductor discharge outlet. (Repeat for all eductors.) Hooks may be installed
on longer tubes to allow discharge tube to conveniently hang from dispenser when not in use.
7. Replace cabinet cover. Push the sides in, behind the latch holes, to snap the cover in place. The two screws provided
may be installed in the holes in the cabinet sides to prevent easy removal of cover.
8. Connect water supply hose of at least 3/8 " ID to water inlet swivel. (Minimum 25 PSI pressure,
with water running, is required for proper operation.) Connect opposite end of hose to water
supply. Turn water supply on.
9. Push button to start ow of desired water/concentrate solution, and hold until supply tube is
primed ( lled). Then push the button whenever dispensing is desired, and release button to
stop ow of solution. If you wish to be able to lock the button in the “on” position: clip or bend
the two tabs behind the lower front portion of the button (see diagram). This allows the button
to be fully depressed and allows it to latch in the “on” position. To unlock, pull the button out.
Troubleshooting Chart:
* In hard water areas, scale may form inside the discharge end of the eductor, as well as in other areas of the unit that are exposed
to water. This scale may be removed by soaking the eductor in a descaling solution (deliming solution). To remove an eductor
located in the cabinet, rmly grasp water valve and unthread eductor. Replace in same manner. Alternatively, a scaled eductor can
be cleaned (or kept from scaling) by drawing the descaling solution through the unit. Operate the unit with the suction tube in the
descaling solution. Operate the unit until solution is drawn consistently, then ush the unit by drawing clear water through it for a
minute. Replace concentrate container and put suction tube into concentrate.
1. No discharge a. No water a. Open water supply
b. Magnetic valve not functioning b. Install valve parts kit
c. Eductor clogged c. Clean* or replace
d. Clogged water inlet strainer d. Disconnect inlet water line and clean s
strainer
2. No concentrate draw a. Clogged foot valve a. Clean or replace
b. Metering tip or eductor has b. Clean (descale)* or replace
scale build-up
c. Low water pressure c. Minimum 25 PSI (with water running)
required to operate unit properly
d. Discharge tube and/or ooding d. Push tube rmly onto eductor discharge
ring not in place hose barb, or replace tube if it doesn't
have a ooding ring.
e. Concentrate container empty e. Replace with full container
f. Inlet hose barb not screwed f. Tighten, but do not overtighten
into eductor tightly
g. Clogged water inlet strainer g. Disconnect inlet water line and clean
strainer
h. Air leak in chemical pick-up tube h. Put clamp on tube or replace tube if brittle
i. Selector out of position i. Assure selector is in position desired
3. Excess concentrate draw a. Metering tip not in place a. Press correct tip rmly into barb on
eductor
b. Chemical above eductor b. Place concentrate below the eductor
4. Failure of unit to turn o a. Water valve parts dirty or a. Clean* or replace with valve parts kit
defective
b. Magnet doesn't fully return b. Make sure magnet moves freely.
c. Push button stuck c. Remove button and clean cabinet/button
to remove any dirt lodged in slide recess
5. Excess foaming in discharge a. Air leak in pick-up tube a. Put clamp on tube or replace tube if brittle
SolutionCause
Problem
THANK YOU FOR YOUR INTEREST IN OUR PRODUCTS
protective clothing and eyewear when dispensing chemicals or other materials.
observe safety and handling instructions of the chemical manufacturers.
direct discharge away from you or other persons or into approved containers.
dispense cleaners and chemicals in accordance with manufacturer's instructions. Exercise CAU-
TION when maintaining your equipment.
equipment clean to maintain proper operation.
protective clothing and eyewear when working in the vicinity of all chemicals, lling or emptying
equipment or changing metering tips.
re-assemble equipment according to instruction procedures. Be sure all components are rmly
screwed or latched into position.
only to tap water outlets (85 PSI maximum).
If the unit is used to ll a sink or the discharge hose can be placed into a sink, the unit must be
mounted so that the bottom of the cabinet is above the over ow rim of the sink.
WEAR
ALWAYS
ALWAYS
ALWAYS
KEEP
WEAR
ALWAYS
AT TACH
Hydro Systems manufactures quality chemical proportioning equiment. Please use this equipment carefully
and observe all warnings and cautions.
********************************************************* NOTE *********************************************
NOTE