Dräger Medical AG & Co. KGaA 6173.3
Babylog 8000 Replacing Non-Repairable Items
2 Cleaning or Replacing the Cooling Air Filter Every 4 Weeks
Pull out the cooling air filter 1 from its mount in the rear panel.
Insert a new cooling air filter or wash the old one in warm water with a few drops of dish-washing
liquid. Dry well.
Insert the cooling air filter 1 into the mount (make sure it is free of folds).
Fig. 1: Cooling Air Filter
1

Table of Contents

Other manuals for Dräger Babylog 8000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Babylog 8000 and is the answer not in the manual?

Dräger Babylog 8000 Specifications

General IconGeneral
Leak CompensationYes
TypeVentilator
Pressure Range0 to 80 mbar
Tidal Volume Range2-2000 mL
Frequency1-150 breaths/min
Inspiratory Time0.1-3.0 sec
PEEP0 to 20 mbar
FiO2 Range21-100%
Power Supply100 to 240 V AC, 50/60 Hz