EasyManuals Logo

Dräger Infinity M540 User Manual

Dräger Infinity M540
610 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #534 background imageLoading...
Page #534 background image
Troubleshooting
534 Instructions for use – Infinity Acute Care System – Monitoring Applications VG6.n
Non-invasive blood pressure (NIBP)
Priority Message Parameter field Problem Solution
!! NIBP S > (alarm limit)
NIBP S < (alarm limit)
Parameter value The parameter
value is
above/below the
upper/lower
alarm limits.
Check the patient
and treat if
necessary.
Change the alarm
limits.
!! NIBP D > (alarm limit)
NIBP D < (alarm limit)
Parameter value
!! NIBP M > (alarm limit)
NIBP M < (alarm limit)
Parameter value
!! %0 H/W failure
1)
Parameter values
are replaced by ***
NIBP
measuremen
t circuit failure
NIBP zero
out of range
or faulty
transducer
Check all hardware,
contact
DrägerService
®
.
1)
%0 is a placeholder for the parameter label NIBP.
!! %0 low inflation
limit
1)
The message Last
measurement
failed! is followed
by the message
Low inflation limit
Parameter values
are replaced by ***
The pressure of
the patient is
greater than the
maximum
allowed cuff
inflation
pressure.
Select the next higher
inflation limit setting.
! NIBP mean only Parameter values
are replaced by ***
The pulse
amplitude is too
small or too high
for the M540 to
derive systolic
and diastolic
pressure values
but sufficient to
report a mean
pressure value.
Check the patient
and treat if
necessary.
Check the hose
and cuff.
Check the size and
the placement of
the cuff.

Table of Contents

Other manuals for Dräger Infinity M540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity M540 and is the answer not in the manual?

Dräger Infinity M540 Specifications

General IconGeneral
BrandDräger
ModelInfinity M540
CategoryMonitor
LanguageEnglish

Related product manuals