Patient Monitor User Manual Basic Operation
- 29 -
shortcut key to restart recording.
NOTE:
The monitor is unable to enter into standby mode when exporting data.
3.3.3 Night Mode
To switch to night mode, you may:
Select the shortcut key on the main screen, or
Select Menu > Common Function > Night Mode.
NOTE:
In night mode, the sound of key, heart beat and pulse is muted; the alarm volume and
screen brightness are down to their minimum; the settings including key volume, beat
volume, PR volume, alarm volume and screen brightness are unavailable.
3.3.4 Privacy Mode
Only if the monitor is connected and admitted by MFM-CMS, the privacy mode can be activated.
To enter into privacy mode, you can select Menu > Maintenance > User Maintain > Shortcut
Setup > Privacy Mode (it is defaulted to be off). Press the shortcut key on the screen, the
monitor enters into privacy mode after user’s confirmation.
In privacy mode:
1. The screen displays message: Privacy mode and Patient is in monitoring without audio
alarm and alarm indicator lighting. Please click screen or hard key to exit.
2. Monitoring data, alarm information, stored data and monitor status are transmitted to
MFM-CMS.
3. Audio alarm paused status discontinues. Audio alarm off, alarm off, alarm reset and alarm
latch status are not influenced.
The monitor exits privacy mode in any of the conditions:
1. The user clicks anywhere on the screen or presses any key (except Power ON/OFF key).
2. Battery Low alarm occurs.
3. The monitor is disconnected with MFM-CMS.
NOTE:
The monitor is unable to enter into privacy mode when exporting data.
3.3.5 NFC Mode
NFC mode is designed when users need to constantly observe HR physiological alarm. In NFC

Table of Contents