EasyManuals Logo

Endress+Hauser Liquiphant M FTL50 Technical Information

Endress+Hauser Liquiphant M FTL50
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Page #1 background imageLoading...
Page #1 background image
TI328F/00/en
Technical Information
Liquiphant M FTL50(H), FTL51(H)
Vibration Limit Switch
Level limit switch for all liquids.
Suitable for use in hazardous areas, food and pharmaceuticals
Application
The Liquiphant M is a level limit switch for use in all
liquids
for process temperatures of –50 °C to 150 °C
for pressures up to 100 bar
for viscosity up to 10000 mm
2
/s
for densities 0.5 g/cm
3
or 0.7 g/cm
3
other settings
available on request
foam detection on request
The function is not affected by flow, turbulence, bubbles,
foam, vibration, bulk solids content or build-up.
The Liquiphant is thus the ideal substitute for float
switches.
FTL50:
Compact design, ideal for mounting in pipes and for
installation in areas difficult to access
FTL51:
With extension pipe up to 3 m
(6 m on request)
FTL50H, FTL51H:
With polished tuning fork and easy-to-clean process
connections and housings for food and pharmaceutical
applications.
High corrosion-resistant AlloyC4 (2.4610) is available for
the fork and process connections for applications in very
aggressive liquids.
EEx ia, EEx de and EEx d protection enable it to be used
in hazardous areas.
Your benefits
Use in safety systems requiring functional safety
to SIL2/SIL3 in accordance with IEC 61508/
IEC 61511-1
Large number of process connections to choose from:
universal usage
Wide variety of electronics, e.g. NAMUR, relay,
thyristor, PFM signal output: the right connection for
every process control system
PROFIBUS PA protocol:
for commissioning and maintenance
No calibration: quick, low-cost start-up
No mechanically moving parts: no maintenance,
no wear, long operating life
Monitoring of fork for damage: guaranteed function
FDA approved materials (PFA Edlon)

Table of Contents

Other manuals for Endress+Hauser Liquiphant M FTL50

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Liquiphant M FTL50 and is the answer not in the manual?

Endress+Hauser Liquiphant M FTL50 Specifications

General IconGeneral
BrandEndress+Hauser
ModelLiquiphant M FTL50
CategorySwitch
LanguageEnglish

Related product manuals