displays 0. Then place the article in container, at this time, it shows the net weight
of the article, remove the article and the container, the scale shows a negative value of
container weight. Still press TARE to return to 0.
3.2.3 LIGHT” backlight key: press “LIGHT” key to choose the backlight mode,
B OFF backlight always off, “B ON backlight always on,“ B AUTObacklight on
when the data changes, Off when the data reaches stabilization after 3 seconds.
3.2.4 COUNT count key: sampling of article by COUNT key to enter the
counting state
(1) Choose the quantity of samples: To precisely count the articles, firstly the quantity
of samples for counting should be chosen in according with the weights of the articles.
The quantity of samples can be chosen from 1-10-20-50-100. For a small number of
articles, the larger quantity of samples, the more precise.
(2) Place one or more samples on the empty pan, the balance will show the total
weights of samples. Then press COUNT, the displaying will show 1, at the same
time, the right of the displaying show unit pcs, with indicate that scale is now in
counting state, and all the samples are one unit. Then press UNIT, the displaying
will show between 1102050100. The quantity of articles and samples must
be the same, and place the similar articles. The total amount of objects is showed,
press the COUNT again back to the normal situation.
3.2.5 UNIT unit conversion key
Press UNIT, the units can be changed between kg. lb”. (1lb=453.59g).
3.2.6 CAL Calibration : Holding on tare key, then press unit until the
displaying window shows C XXX and the scale will automatic calibrate,( XXX
should be the weight of standard weights, for example: when displaying showing C
20 express the need to place a standard weight 20kg), at the same time, only place
calibrating weights on the platform, after stability, the scale displays the weight of the
calibrating weights and the unit kg, till now the calibration is completed and the
scale can be normally used.