DINAMAP non-invasive blood pressure
Technology DINAMAP
®
classic and SuperSTAT
™
(SuperSTAT only
available with Masimo and Nellcor SpO
2
)
Measurement technique Oscillometric
Displayed parameters Systolic, diastolic and mean pressures, time of
last measurement
Measurement modes Adult ICU and OR modes; manual, auto and stat,
neonatal mode; manual and auto
Systolic
Adult 30-285 mmHg
Pediatric 30-235 mmHg
Neonate 30-140 mmHg
Map
Adult 20-260 mmHg
Pediatric 20-220 mmHg
Neonate 20-125 mmHg
Diastolic
Adult 10-220 mmHg
Pediatric 10-210 mmHg
Neonate 10-110 mmHg
Pulse rate, as displayed in tabular trends or 3 wave form display
Adult 30-200 bpm
Pediatric 30-200 bpm
Neonate 30-200 bpm
Other specifications
Overall system accuracy Meets or exceeds SP 10-1992 AAMI standards
Automatic cycle times 0-4 hours
Tubing length 12 feet adult, 8 feet neonatal
Automatic cuff deflation Cycle time exceeding 3 minutes (90 seconds neonatal),
French mode – Cycle time exceeding 2 minutes (60 seconds
neonatal), power off, or cuff pressure exceeds 294 mmHg
(± 6 mmHg) adult, 250 (± 5 mmHg) pediatric,
147 (± 3 mmHg) neonatal
Cuff sizes Thigh, large adult, adult, small adult, child, infant and
neonatal, sizes #5 - #1 and assorted long sizes
Alarms User-selectable upper and lower limits for systolic,
diastolic and mean pressures