Care and Maintenance
4-28 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
off open (reversed
polarity)
500
on closed
(reversed
polarity)
100
on open (reversed
polarity)
500
Table 4-11: Typical data format for recording enclosure/chassis leakage (Continued)
Unit under test____________________________________ Date of test:_____________
Test Conditions Measurement/Test Point Location
System
Power
Grounding/
PE
Limit
µA
Potential
equilibrium
connector
Monitor
housing
Probe
connector