IRECTION 5813707-100, REVISION 2 VENUE GO™ SERVICE MANUAL
Chapter 4 - General Procedures and Functional Checks 4-5
PRELIMINARY
4-2-3 Power Shutdown
When you tap the on/off button to turnoff the system, the system displays the following options:
Figure 4-46 Exit Dialog Window
Standby
Use this button to enter the system into Standby mode.
Exit
(Only available when logged in as GE Service with Service Dongle)
Select this button when you want to exit to the Windows Desktop.
NOTE: If you need to restart Venue Go™ when logged on to the Windows Desktop, ensure that you do a
complete power down (Shut Down). This is required to power up the Front End Processor.
Shutdown
Use this button to shut down the system. The entire system will shut down.
If the Shutdown button is greyed out, use the key-combination <Ctrl+Alt+R> to shut down the unit.
•Cancel
Use this button to exit from the System-Exit menu and return to the previous operation.
4-2-4 Logging On to the Venue Go™ as “ADM”
4-2-4-1 Open the Config screen
1) Select Settings.
2) Select Config
This brings up the Operator Login dialog where you can log on to Venue Go™ .