IRECTION 5813707-100, REVISION 2 VENUE GO™ SERVICE MANUAL
Chapter 4 - General Procedures and Functional Checks 4-5
PRELIMINARY
4-2-3 Power Shutdown
When you tap the on/off button to turnoff the system, the system displays the following options:
Figure 4-46 Exit Dialog Window
• Standby
Use this button to enter the system into Standby mode.
• Exit
(Only available when logged in as GE Service with Service Dongle)
Select this button when you want to exit to the Windows Desktop.
NOTE: If you need to restart Venue Go™ when logged on to the Windows Desktop, ensure that you do a
complete power down (Shut Down). This is required to power up the Front End Processor.
• Shutdown
Use this button to shut down the system. The entire system will shut down.
If the Shutdown button is greyed out, use the key-combination <Ctrl+Alt+R> to shut down the unit.
•Cancel
Use this button to exit from the System-Exit menu and return to the previous operation.
4-2-4 Logging On to the Venue Go™ as “ADM”
4-2-4-1 Open the Config screen
1) Select Settings.
2) Select Config
This brings up the Operator Login dialog where you can log on to Venue Go™ .