IRECTION 5813707-100, REVISION 2 VENUE™ SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-3
PRELIMINARY
8-2-2 Batteries Replacement Procedure
8-2-2-1 Tools
N/A
8-2-2-2 Time Required
5 min
8-2-2-3 Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in Power Shutdown.
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in Mounting / Dismounting System on Cart
and place it on a previously prepared surface with the display facing down.
4.) Open the battery switch silicon cover and switch off the battery power.
Figure 8-71 Battery Power Switch