10-18 Section 10-8 - Electrical Safety Tests
Table 10-22 Total Patient leakage current: Type CF probe to Mains (Only for Europe)
NOTE: * Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards Table IV
** Measurement limits per UL 60601-1 Medical Electrical Equipment Safety Standards, Table
19.5DV.1 *** Measurement limits per IEC 60601-1 Ed 2. Medical Electrical Equipment Safety
Standards, Table IV
**** Isolation between Mains and Applied Limits refers to the measurement of leakage current
flow which would flow from mains to patient if the patient were to come into contact with mains
voltage.
*****Measurement limits per IEC 60601-1 Ed 3. and ANSI/AAMI ES60601-1, Medical Electrical
Equipment Safety Standards, Table 3
******Measurement limits per IEC 60601-1 Ed 3.and ANSI/AAMI ES60601-1, Medical Electrical
Equipment Safety Standards, Table 4
The following tests are performed at the factory and should be performed at the customer site as well.
- Grounding Continuity
- Earth Leakage Current
- Enclosure Leakage Current
- Patient Leakage Current
All measurements should be made with an electrical safety analyzer.
Conditions Limitation
Close Protective earth and close neutral with normal polarity 0.1 mA******
Close Protective earth and close neutral with reverse polarity 0.1 mA******

Table of Contents

Other manuals for GE Voluson P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Voluson P6 and is the answer not in the manual?

GE Voluson P6 Specifications

General IconGeneral
BrandGE
ModelVoluson P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals