4. Select a 4D Biopsy user setting (e.g., Default).
The preset values are loaded.
5. Select the desired “Biopsy Mode”:2D image + volume box are shown on the screen. (no
biopsy line)
2D image + biopsy line + volume box are shown on the screen.
2D image + volume box are shown on the screen. (no biopsy line)
6. Place the Volume box over the region of interest.
The trackball has two functions: position and size of the Volume box. The activated function
is displayed in status bar area on the monitor.
Press the upper trackball key to change the function of the trackball from position to size or
vice versa.
Volume Mode
9-114
Voluson® S6/S8 Basic User Manual
5433669-100 Revision 4