Chapter 1 31
General Information
HP 8529B Test and Adjustment Module
HP 8529B Test and Adjustment Module
When attached to the spectrum analyzer rear panel, the HP 85629B
Test and Adjustment Module (TAM) provides diagnostic functions for
the HP 8560E/EC. Because the TAM connects directly to the spectrum
analyzer internal data and address bus, it controls the spectrum
analyzer hardware directly. It would be impossible to control the
hardware to the same extent either from the spectrum analyzer front
panel or over the HP-IB.
The TAM measures voltages at key points in the circuitry and flags a
failure whenever the voltage falls outside the limits. The TAM locates
the failure to a small functional area which can be examined manually.