Function and Features
Copyright © Itech Electronics Co., Ltd. 27
convenient. Customer can save and recall parameters quickly.
GRP1: This indicates saving power supply parameters in 1-9 groups.Press
(Shift)+ (Save) and the group number(1-9) can save the parameters
in corresponding groups.
GRP2: This indicates saving the parameters in 10-18 groups.Press (Shift) +
(Save)+saved group number(1-9)can save related parameters.Note
that the current number “1” represents parameters are saved in 10
th
groups.Number “2” represents the parameters are saved in 11
th
groups.
GRP3-GRP8 by parity of reasoning.
Detailed Save and Recall operation refer to chapter 3.7.
NOTE
IT6726G/IT6726H has 9×7 groups of memory capacities.
Timer Set (TIMER SET)
This item is used to set the “time on- load” function, time range 0.1-99999S .In
ON mode, the indicator light “Timer” will be lit on the VFD screen.When output
of power supply is opened, timer will begin to work, after reaching the definite
time,output will be off automatically.If in OFF mode,the timer function will not be
enabled.The default set is OFF.
Reset (RESET)
This item is used to reset all items in the menu.If you select YES, then unit will
restored to factory setting.If you select NO, all settings in the menu will remain
unchanged.
List (List Set)
IT6700 series power supply provides 9 list files, each list file includes 150 steps.
Before you edit a list file, please set the trigger mode:manual mode.
Press (Shift)+ (Menu)button to enter the menu,then press direction
key to select >SYST SET option,after that please push button to
confirm.At last to press direction key to select >Trig MANUAL and push
button to confirm.
You can make the output change order by editing every step value of list
operation. The parameters you need to edit includes:single-step
voltage,single-step current,single-step delay time and whether to go on the next
step.Besides,you also need to set the repeat times and save list sequence
file.After the editing process,at this time if a trigger signal is received,power
supply will begin to work according to the sequence steps you’ve edit. Now we
take five steps for an example.
Operation steps:
1. Press (Shift) + (Menu)button to enter the menu.
2. VFD display >MAX VOLT,press to select >LIST SET,press
to confirm
3. VFD display >LIST STATE,press to select >LIST EDIT, press
to confirm.