18-8 Performance Verification
Performing the verification test procedures
Test summary
•Volts measurement accuracy
Amps measurement accuracy
Ohms measurement accuracy
Coulombs measurement accuracy
If Model 6514 is not within specifications and not under warranty, see the calibration proce-
dures in Section 19 for information on calibrating the unit.
Test considerations
When performing the verification procedures:
Restore Model 6514 factory front panel defaults and perform input bias current and volt-
age offset calibration as outlined below.
Make sure that the test equipment is properly warmed up and properly connected to
Model 6514 INPUT jack.
Be sure test equipment is set up for the proper function and range.
Allow the input signal to settle before making a measurement.
Do not connect test equipment to Model 6514 through a scanner, multiplexer, or other
switching equipment.
WARNING The maximum common-mode voltage (voltage between analog common
and chassis ground) is 500V peak, DC to 60Hz sine wave. Exceeding this
value may cause a breakdown in insulation, creating a shock hazard.
CAUTION The maximum input voltage is 250V peak, DC to 60Hz sine wave. Exceed-
ing this voltage may result in instrument damage.

Table of Contents

Other manuals for Keithley 6514

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6514 and is the answer not in the manual?

Keithley 6514 Specifications

General IconGeneral
BrandKeithley
Model6514
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals