Kodiak Recirculating Chiller Technical Manual
Manual # 820-0109, Rev. Y, 07/08/10 - 22 - Check Lytron’s website www.Lytron.com
for the most current technical information.
AL—Audible Alarm (option)
1. To enter the menu, press and hold both the “UP” and “DOWN” arrow keys for 3 seconds.
2. Use the “SET” key to scroll to “AL.”
3. The display toggles between “AL” and its current setting.
4. Use “UP” or "DOWN" arrow key to toggle between “ON” and “OFF.”
5. Press, “SET” to accept the setting on the display and continue to the next feature.
Ar - Auto Restart Feature (standard)
If auto restart is enabled (ON):
If the chiller is running and the power is interrupted due to power outage, tripped circuit breaker, etc.,
the chiller will automatically start when the power is restored.
If auto restart is enabled (OFF):
If the chiller is running and the power is interrupted due to power outage, tripped circuit breaker, etc.,
the chiller will stay off when the power is restored. Push “ON” to restart the chiller.
To activate the auto restart feature:
1. Enter the customer menu by holding both the “UP” and “DOWN” arrow keys for 3 seconds.
2. Use the “SET” key to scroll to “Ar.”
3. The display toggles between “Ar” and its current setting.
4. Use “UP” or “DOWN” arrow key to toggle between “ON” and “OFF.”.
5. Press, “SET” to accept the setting on the display and continue to the next feature.
CAL - Calibration offset (standard)
1. To enter the menu, press and hold both the “UP” and “DOWN” arrow keys for 3 seconds and release.
2. Use the SET key to scroll to “CAL.”
3. The display toggles between “CAL” and current offset setting.
4. Use “UP” or “DOWN” arrows to change the calibration offset, which must be between -4°C and +4°C. If
you do not want to use an offset set it to zero (0).
5. Press, “SET” to accept the setting on the display and exit the menu.
DEGC - Temperature scale (standard)
1. To enter the menu, press and hold both the “UP” and “DOWN” arrow keys for 3 seconds and release.
2. The first option to change will be “DEGC.”
3. The display toggles between “DEGC” and current setting. Use “UP” or “DOWN” arrows to turn the
degree Celsius display on and off. This will automatically change the pressure scale to psi or to bar.
4. Press, “SET” to accept the setting on the display and exit the menu.
FS—Fault Chiller Shutoff (option)
1. To enter the menu, press and hold both the “UP” and “DOWN” arrow keys for 3 seconds.
2. Use the “SET” key to scroll to “FS.”
3. The display toggles between “FS” and its current setting.
4. Use “UP” or "DOWN" arrow key to toggle between “ON” and “OFF.”
5. Press, “SET” to accept the setting on the display and continue to the next feature.
Lt— Low temperature indication (option)
1. To enter the menu, press and hold both the “UP” and “DOWN” arrow keys for 3 seconds.
2. Use the “SET” key to scroll to “Lt.”
3. The display toggles between “Lt” and current over-temperature set point
4. Use “UP” or “DOWN” arrows to change the low temperature set point, which should be at least 3°C
(6°F) below the coolant set point temperature.
5. Press, “SET” to accept the setting on the display and continue to the next feature.