Time Wei
g
htin
g
Default to " Fast " time wei
g
htin
g
characteristics.
* " Fast time wei
g
htin
g
" is simulated the
human ear response character.
Accuracy Characteristics of " A " frequency
(
23± 5
)
wei
g
htin
g
network meet IEC 61672
class 2.
Under 94 dB input si
g
nal, the accuracy
are :
31.5 Hz ± 3.5 dB
63 Hz ± 2.5 dB
125 Hz ± 2.0 dB
250 Hz ± 1.9 dB
500 Hz ± 1.9 dB
1
K
Hz ± 1.4 dB
2
K
Hz ± 2.6 dB
4
K
Hz ± 3.6 dB
8
K
Hz ± 5.6 dB
Remark :
T
he above spec. are tested under th
e
environment RF Field Strength less tha
n
3 V/M & frequency less than 30 MH
z
only.
Calibrator B & K (Bruel & k
j
aer), multi-fuction
acoustic calibrator, model : 4226.
Microphone Electric condenser microphone.
Size of microphone 1/2 inch standard size.
Calibration VR Build in external calibration VR, easy to
calibrate on 94 dB level by screw driver.
Data Hold Hold function to freeze the display value.
Operatin
g
Temp. 0 to 50 ( 32 to 122 ).℃℉
Operatin
g
Humidity Less than 80% RH.
Power Supply
Alkaline or Heavy duty type, 9V d.c.
006 P, MN1604(PP3) or equivalent.
Power Consumption Approx. DC 6 mA.
Dimension 250 x 70 x 28 mm ( 9.9 x 2.8 x 1.1 inch ).
Wei
g
ht 250
g
/0.55 LB ( includin
g
battery ).
Standard Instruction Manual ...................... 1 PC.
Accessories
Optional 94 dB Sound Calibrator,
Accessories Model : SC-941, SC-942.
Hard carryin
g
case, Model : CA-06.
2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Lutron Electronics SL-4010 and is the answer not in the manual?

Lutron Electronics SL-4010 Specifications

General IconGeneral
BrandLutron Electronics
ModelSL-4010
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals