Amicon
®
Ultra-4 Centrifugal Filter Devices
Introduction
Amicon
®
Ultra-4 centrifugal lter devices provide fast ultraltration, with the capability for high concentration factors and easy concentrate
recovery from dilute and complex sample matrices. The vertical design and available membrane surface area provide fast sample
processing, high sample recovery (typically greater than 90% of dilute starting solution), and the capability for 80-fold concentration.
Typical processing time is 10 to 40 minutes depending on Molecular Weight Cuto (MWCO). Solute polarization and subsequent fouling
of the membrane are minimized by the vertical design, and a physical deadstop in the lter device prevents spinning to dryness and
potential sample loss. The concentrate is collected from the lter device sample reservoir using a pipettor, while the ultraltrate is
collected in the provided centrifuge tube. The device can be spun in a swinging-bucket (for optimal performance) or xed-angle rotor.
Amicon
®
Ultra-4 devices are supplied nonsterile and are for single use only.
The Amicon
®
Ultra-4 product line includes 5 dierent cutos (MWCO):
Amicon
®
Ultra 3K device — 3,000 MWCO
Amicon
®
Ultra 10K device — 10,000 MWCO
Amicon
®
Ultra 30K device — 30,000 MWCO
Amicon
®
Ultra 50K device — 50,000 MWCO
Amicon
®
Ultra 100K device — 100,000 MWCO
Applications
Concentration of biological samples containing antigens, antibodies, enzymes, nucleic acids (DNA/RNA samples, either single- or
double-stranded), microorganisms, column eluates, and puried samples
Purication of macromolecular components found in tissue culture extracts and cell lysates, removal of primer, linkers, or molecular
labels from a reaction mix, and protein removal prior to HPLC
Desalting, buer exchange, or dialtration
Materials Supplied
The Amicon
®
Ultra-4 device is supplied with a cap, a lter device, and a centrifuge tube.
Cap
Filter device
Centrifuge tube
Required Equipment
Centrifuge with swinging-bucket (preferred) or xed-angle rotor with wells/carriers that can accommodate 17 mm × 124 mm 15 mL
conical-bottomed tubes
CAUTION: To avoid damage to the device during centrifugation, check clearance before spinning.
Pipettor with 200 microliter (μL) tip for concentrate recovery
User Guide
Amicon
®
Ultra-4 Centrifugal Filter Devices
for volumes up to 4 mL
Amicon
®
Ultra-4 3K, 10K, 30K, 50K, and 100K devices for research use only; not for use in diagnostic procedures

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Millipore Amicon Ultra-4 Series and is the answer not in the manual?

Millipore Amicon Ultra-4 Series Specifications

General IconGeneral
BrandMillipore
ModelAmicon Ultra-4 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals