Page #1 background image
OLYMPUS KEYMED
KeyMed House, Stock Road, Southend-on-Sea,
Essex SS2 5QH, United Kingdom
Telephone: +44 (0)1702 616333 Fax: +44 (0)1702 465677
e-mail: keymed@keymed.co.uk www.keymed.co.uk
For a complete listing of
sales and distribution locations visit:
www.olympus.com
SUCTION PUMP
KV-6
Specification
Classification
(electromedical
equipment)
Standards compliance Complies with EN ISO 10079-1 and EN IEC 60601-1 and UL 60601-1
Electro-magnetic
Compatibility
This product complies with the requirements of EN IEC 60601-1-2 for emissions and immunity, and as such, its operation is unlikely to
be affected by, or cause interference with, equipment meeting appropriate EMC standards. As a precaution, equipment which may be
sensitive to interference outside the limits specified by EN IEC 60601-1-2 should not be placed in close proximity to the KV-6.
Type of protection against
electrical shock
Classification according to EN IEC 60601-1 / UL 60601-1: Class 1 with Type CF applied part.
Degree of protection
against electrical shock
In accordance with EN IEC 60601-1 and UL 60601-1, the KV-6 is marked with the symbol to indicate the provision of an adequate
degree of protection against electrical shock and that it has an applied part isolated from all other parts of the equipment.
Degree of protection
against explosion
None: the Olympus KV-6 must NOT be used in the zone of risk of flammable anaesthetic gases.
Mode of operation Continuous
Regulatory status European Economic Area
(EEA)
CE This mark on the product indicates compliance with Directive 93/42/EEC relating to medical devices, Class IIa.
Dimensions and weight Dimensions Height: 210mm Width: 305mm Depth: 375mm
Weight 9.3 kg
Power switch Marking The power switch is marked . A white LED in the center signalizes it is turned on
Power requirements Power supply 100-240 V~ Fusing 2 x T1.6 AH Marking: The mark ~ on
the product indicates the
requirement for an AC
power supply
Frequency 50/60Hz Power rating 120 W
Environmental
conditions
Ambient Temperature Operational: +5ºC to +40ºC (+41ºF to +104ºF)
Storage: -20ºC to +50ºC (-4ºF to +122ºF)
Relative humidity Operational: 30-75 %
Storage: 20-95%
Atmospheric pressure Operational/Storage: 70.0 to 106 kPa
Fluid Ingress In accordance with EN IEC 60601-1/UL 60601-1, the product is marked IP21 to indicate that it is provided with an enclosure that prevents entry of such an amount of
falling liquid as might interfere with its safe and satisfactory operation when correctly positioned.
Resistance to
chemicals
The external surfaces of the Olympus KV-6 are resistant to: 2% aqueous neutral detergent, 70% ethyl alcohol, isopropyl alcohol, water.
Pump specification Nominal vacuum: 95 kPa Nominal free air flow rate: 50 l/min, adjustable to 40 or 60 l/min.
In accordance with ISO 10079-1, the unit is marked with the words ‘HIGH VACUUM, HIGH FLOW’ to indicate it has a vacuum of at least 60 kPa in 10 seconds and a
free air flow rate of 20l/min or greater.
Thermal cut-out The temperature is observed by a sensor in the housing. If the temperature in the housing exceeds 80 °C (176 °F), the software stops the motor and turns on the yellow
LED. It is possible to restart the motor again as soon as the temperature in the housing drops below 70 °C (158 °F).
Single use suction liner To prevent fluid from being drawn up to the pump the single use suction liner contains a mechanism to shut off the flow when the liner is overfilled. This mechanism is
activated when contact is made with aspirated fluid.
Suction jar The suction jars can be autoclaved up to 135 ºC or disinfected. The jar is impact resistant.
Suction jar lid The suction jar lid clips and where applicable the float mechanism, are autoclavable up to 135 ºC. The inlet connection (from the endoscope) is marked with the word
PATIENT, the outlet connection (to the pump) is marked with the word VACUUM.
Tubing Use only tubing specified in the KV-6 Instructions for Use, otherwise equipment malfunction may result.
Vacuum gauge Range: 0 to 100 kPa Accuracy: ±2.5%
Printed in UK PL022/201505