Thank you for purchasing the OMRON HEM-6131 Wrist Blood
Pressure Monitor.
The OMRON HEM-6131 is a compact and easy to use blood
pressure monitor, operating on the oscillometric principle. It
measures your blood pressure and pulse rate simply and quickly.
For comfortable controlled inflation without the need of pressure
pre-setting or re-inflation the device uses its advanced “IntelliSense”
technology.
Intended Use
This product is designed to measure the blood pressure and pulse
rate of people within the range of the designated wrist cuff, following
the instructions in this instruction manual.
It is mainly designed for general household use. Please read the
Important Safety Information in this instruction manual before using
the unit.
Please read this instruction manual thoroughly before using the
unit.
Please keep for future reference.
For specific information about your own blood pressure,
CONSULT YOUR DOCTOR.
Consult your doctor prior to using in pregnancy or if diagnosed with
arrhythmia or arteriosclerosis.
Please read this section carefully before using the unit.
Warnin g:
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results
and self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should
consult a doctor before using the unit, as cuff inflation can cause
internal bleeding.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Consult a doctor immediately.
Caution:
Indicates a potentially hazardous situation which, if not avoided, may
result in minor or moderate injury to the user or patient or damage to
the equipment or other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who
cannot express their consent.
• Do not use the unit for any purpose other than measuring blood
pressure.
• Do not use a mobile phone or other devices that emit
electromagnetic fields, near the unit. This may result in incorrect
operation of the unit.
• Do not disassemble the unit or wrist cuff.
• Do not operate the unit in a moving vehicle (car, airplane).
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse
with plenty of clean water.
• Use only two “AAA” alkaline (LR03) batteries with this unit.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace both
batteries at the same time.
• Remove the batteries if the unit will not be used for three months or
more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit.
• Do not take measurements after bathing, drinking alcohol or
caffeine, smoking, exercising or eating.
• Do not inflate the wrist cuff when it is not wrapped around your
wrist.
• Read and follow the “Important information regarding Electro
Magnetic Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the
Technical Data Section when disposing of the device and any used
accessories or optional parts.
Main unit:
Display:
2.1 Installing/Replacing the Batteries
• If the battery low symbol ( ) appears on the display, turn off
the unit then replace both batteries at the same time.
• The measurement values continue to be stored in memory
even after the batteries are replaced.
Disposal of used batteries should be carried out in accordance
with the national regulations for the disposal of batteries.
2.2 Setting the Date and Time
Notes:
• If the batteries have been removed, the date and time setting
will need to be reset.
• If the date and time are not set, “-:--” appears during or after
measurement.
For more information about re-setting the date and time,
please refer to 3.4 Setting the Positioning Sensor.
3.1 Applying the Wrist Cuff
Do not apply over clothing.
3.2 How to Sit Correctly
• Relax your wrist and hand. Do not bend your wrist back, clench
your fist, or bend your wrist forward.
3.3 Taking a Reading
Notes:
• To cancel a measurement, press the START/STOP button at
any time during measurement.
• Remain still while taking a measurement.
After the Positioning symbol lights for more than 2 seconds,
the wrist cuff will start to inflate automatically.
Cuff Wrapping Guide
The Cuff wrapping Guide is a unique feature that indicates if
the cuff is not wrapped tightly enough around the wrist.
If the is displayed, it means the reading is NOT reliable
due to the incorrect wrapping of the cuff.
Please wrap the cuff again, taking care to wrap it correctly and
take the measurement again. When the is displayed,
the cuff is correctly wrapped tightly enough on the wrist and
the reading is accurate and reliable.
The monitor automatically stores the measurement in its
memory.
It will automatically turn off after two minutes.
This criteria is for home blood pressure measurement.
If the irregular heartbeats cause the measurement to be
invalid, no result is shown. If the irregular heartbeat symbol
( ) is shown after you have taken a measurement, repeat
the measurement. If the irregular heartbeat symbol ( ) is
shown frequently, please make your doctor aware of it.
Notes:
• Wait 2 - 3 minutes before taking another blood
pressure measurement. Waiting between readings
allows the arteries to return to the condition prior to
taking the blood pressure measurement.
• The blood pressure can differ between the right arm
and the left arm, and therefore also the measured
blood pressure values can be different. OMRON
recommends to always use the same arm for
measurement. If the values between the two arms
differ substantially, please check with your doctor
which arm to use for your measurement.
Warning:
Self-diagnosis of measured results and treatment are
dangerous. Please follow the instructions of your doctor.
3.4 Setting the Positioning Sensor
For more information about setting the date and time, please
refer to 2.2 Setting the Date and Time.
3.5 Using the Memory Function
The unit automatically stores up to 60 sets of measurement
values (blood pressure and pulse rate).
It can also calculate an average reading based on the
measurements from the last three readings taken within 10
minutes. If there are only two readings in memory for that
period, the average will be based on two readings. If there is
one reading in memory for that period, the average will be
based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest
readings.
• When viewing the reading taken without setting the date and
time, “-:--” is displayed instead of the date and time.
To View the Average Value
To View the Readings Stored in Memory
To Delete All the Values Stored in Memory
Note: You cannot partially delete the stored readings.
4.1 The Icons and Error Messages
Note: The irregular heartbeat symbol ( ) and the movement error
symbol ( ) may also be displayed with error messages.
4.2 Troubleshooting
4.3 Maintenance
To protect your unit from damage, please observe the
following:
• Do not subject the main unit and the cuff to
extreme temperatures, humidity, moisture or
direct sunlight.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or
vibrations (for example, dropping the unit on the
floor).
• Do not use volatile liquids to clean the main unit.
• Do not wash the cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to
clean the cuff.
• Do not carry out repairs of any kind yourself. If a
defect occurs, consult your OMRON retail outlet
or distributor as mentioned on the packaging.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean
the cuff.
• Keep the unit in its storage case when not in
use.
• Fold the cuff into the storage case.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures,
humidity, direct sunlight, dust or corrosive
vapours.
• Locations exposed to vibrations, shocks or
where it will be at an angle.
Calibration and Service
• The accuracy of this blood pressure monitor has
been carefully tested and is designed for a long
service life.
• It is generally recommended to have the unit
inspected every two years to ensure correct
functioning and accuracy. Please consult your
authorised OMRON dealer or the local
representative.
Note: Subject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC
(Medical Device Directive).
• This blood pressure monitor is designed according to the European
Standard EN1060, Non-invasive sphygmomanometers Part 1:
General Requirements and Part 3: Supplementary requirements
for electromechanical blood pressure measuring systems.
This OMRON product is manufactured under the strict quality
system of OMRONHEALTHCARE Co., Ltd., Japan. The Core
component for OMRON blood pressure monitors, which is the
Pressure Sensor, is produced in Japan for assembly.
Made in Vietnam
A. Wrist cuff
D. Memory ( ) button
B. Battery compartment
E. Date/Time setting ( ) button
C. Display F. START/STOP button
To take a measurement, you need
to be relaxed and comfortably
seated, under comfortable room
temperature. No bathing, drinking
alcohol or caffeine, smoking,
exercising or eating 30 minutes
before taking a measurement.
• Sit on a chair with your feet flat on the
floor.
• Sit upright with your back straight.
• The cuff should be at the same level
as your heart.
1.
Press the START/STOP button.
All the symbols appear on the display.
While your position is improper, the Positioning symbol and the
Positioning indicator flash, when you are in proper position the
Positioning symbol lights on the display.
The monitor is in
proper position.
The monitor is higher
than the heart.
The monitor is lower
than the heart.
2.
Undo the wrist cuff and remove the unit.
3.
Press the START/STOP button to turn off the
monitor.
Important:
• Recent research suggests that the
following values can be used as a
guide to high blood pressure for
measurements taken at home.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
• Your blood pressure monitor includes an
irregular heartbeat feature. Irregular
heartbeats can influence the results of the
measurement. The irregular heartbeat
algorithm automatically determines if the
measurement is usable or needs to be
repeated. If the measurement results are
affected by irregular heartbeats but the result
is valid, the result is shown together with the
irregular heartbeat symbol ( ).
• If you move during measurement, the
movement error symbol ( ) will appear on
the display. Keep still and repeat the
measurement.
Important information regarding Electro Magnetic
Compatibility (EMC)
With the increased number of electronic devices such as PC’s and
mobile (cellular) telephones, medical devices in use may be
susceptible to electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of
the medical device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic
Compatibility) with the aim to prevent unsafe product situations, the
EN60601-1-2:2007 standard has been implemented. This standard
defines the levels of immunity to electromagnetic interferences as
well as maximum levels of electromagnetic emissions for medical
devices.
This medical device manufactured by OMRON HEALTHCARE
conforms to this EN60601-1-2:2007 standard for both immunity and
emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which
generate strong electrical or electromagnetic fields, near the
medical device. This may result in incorrect operation of the unit
and create a potentially unsafe situation. Recommendation is to
keep a minimum distance of 7 m. Verify correct operation of the
device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is
available at OMRON HEALTHCARE EUROPE at the address
mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its
literature, indicates that it should not be
disposed of, with other household wastes at the
end of its working life.
To prevent possible harm to the environment or human health from
uncontrolled waste disposal, please separate this product from
other types of wastes and recycle it responsibly to promote the
sustainable reuse of material resources.
Household users should contact either the retailer where they
purchased this product, or their local government office, for details
of where and how they can return this item for environmentally safe
recycling.
Business users should contact their supplier and check the terms
and conditions of the purchase contract. This product should not be
mixed with other commercial wastes for disposal.
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto,
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR H oofddorp
THE NETHERLANDS
Asia Pacific HQ
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road , #05-05/08
Alexandra Technopark, Singapore 119967
www.omron-healthcare.com.sg
Production Facility
OMRON HEALTHCARE MANUFACTURING
VIETNAM CO., LTD.
Binh Duong Province, Vietna m
13A0092
HEM-6131-AP_A_M_EN.fm 1 ページ 2013年5月14日 火 曜日 午後6時9分