EasyManuals Logo
Home>Tektronix>Test Equipment>475

Tektronix 475 User Manual

Tektronix 475
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #54 background imageLoading...
Page #54 background image
c. Adjust the A LEVEL control for a stable display.
Calibration475
e. CHECK-Using the CH 2 VOLTS/DIV switch and the
Standard Amplitude Calibrator settings given in Table 5-2,
check that the vertical deflection factor is within the given
tolerance in each position of the CH 2 VOLTS/DIV switch. d. CHECKDuration of each cycle is about 1 division.
9. Check Channel 1 and 2 Variable Volts/Division
Range
a. Set both VOLTS/DIV switches to 20 mV.
b. Adjust the Standard Amplitude Calibrator for a 0.1
volt output.
c. Rotate the CH 2 VAR control fully counter
clockwise.
d. CHECKCRT display reduces to less than 2
divisions.
e. Set the CH 1 AC-GND-DC switch to DC and the
VERT MODE switch to CH 1.
f. Rotate the CH 1 VAR control fully counterclockwise.
g. CHECKCRT display reduces to less than 2 divisions.
h. Return both VAR controls to the detent positions.
i. Disconnect the test setup.
10. Check Alternate Mode Operation
a. Set the VERT MODE switch to ALT and the A
LEVEL control fully clockwise.
b. Position the traces 2 divisions apart.
c. CHECKThat the sweeps alternate in all settings of
the T IM E /D IV switch except X-Y.
11. Check Chop Mode Operation
a. Set the A TIM E /D IV switch to 1 jus, the A SOURCE
switch to NORM, and the VERT MODE switch to CHOP.
e. CHECKCRT display for complete blanking of
switching transients between chopped segments (see
Fig. 5-1).
12. Check Vertical Amplifier Bandwidth
a. Set the TRIG MODE switch to AUTO, the A
TIM E /D IV switch to 0.2 ms, and both VOLTS/DIV
switches to 5 mV.
b. Connect the output of the High-Frequency Constant-
Amplitude Signal Generator (067-0532-01) to the CH 1
input via a GR-to-BNC adapter, X I0 BNC attenuator, and a
50 £2 BNC termination.
c. Adjust the High-Frequency signal generator output
amplitude for a five-division display of a 3 megahertz
reference signal.
d. Without changing the output amplitude, increase the
output frequency of the signal generator until the display is
reduced to 3.5 divisions.
e. CHECKOutput frequency of the High-Frequency
signal generator must be at least 200 megahertz.
f. Move the test signal to the CH 2 input and set the
VERT MODE switch to CH 2.
b. Position the two traces about 4 divisions apart.
Fig. 5-1. Typical CRT display when checking chopped blanking.

Other manuals for Tektronix 475

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 475 and is the answer not in the manual?

Tektronix 475 Specifications

General IconGeneral
BrandTektronix
Model475
CategoryTest Equipment
LanguageEnglish

Related product manuals