EasyManuals Logo
Home>Tektronix>Test Equipment>475

Tektronix 475 User Manual

Tektronix 475
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #74 background imageLoading...
Page #74 background image
Calibration475
e. INTERACTIONBetween Geometry and Y Axis
Alignment adjustments. Adjust both for optimum response.
9. Adjust Z-Axis Compensation
a. Set A T IM E /D IV switch to 0.05 jus.
b. Connect the X I0 probe from the Test Oscilloscope to
TP1366 (see Fig. 5-4).
c. Connect the A +GATE output to the External Trigger
input of the Test Oscilloscope via a 42-inch 50 £2 BNC
cable.
d. Adjust the Test Oscilloscope trigger controls to view
the positive step of a 1 vplt display of the Z-Axis signal.
e. CHECKTest Oscilloscope display for optimum
square corner on the unblanking gate.
f. ADJUSTZ-Axis Compensation adjustment. Cl 352,
using a low capacitance screwdriver, (see Fig. 5-4) for
optimum square corner on the unblanking gate.
g. Disconnect the test setup.
VERTICAL SYSTEM CALIBRATION
Equipment Required
1. Standard Amplitude Calibrator (067-0502-01)
12.
GR-to-BNC Female Adapter
2. Test Oscilloscope
13. 42-inch 50 £2 BNC Cable (two)
3. High-Frequency Constant-Amplitude Signal Generator
14. 18-inch 50 £2 BNC Cable
(067-0532-01)
15.
X I0 BNC Attenuator
4. Medium-Frequency Constant-Amplitude Signal Genera
tor (Type 191)
16.
X I0 GR Attenuator (two)
5. Fast-Rise High-Amplitude Pulse Generator (Type 109)
17.
X5 GR Attenuator
6. Square-Wave Generator (Type 106)
18.
50 £2 BNC Termination (two)
7. Precision DC Voltmeter
19.
GR 50 £2 Termination
8. X I0 Probe (two)
20. 20 Picofarad Normal izer
9. Signal Insertion Unit (067-0676-00)
21.
Dual Input Coupler
10. 50 £2 Signal Pickoff Unit (Type CT-3)
22. Three-inch Screwdriver
11. GR 5 Nanosecond Cable
23.
Low-Capacitance Screwdriver
Control Settings
Preset instrument controls to the settings given under
Preliminary Control Settings except as follows:
VOLTS/DIV 0.1 V
10. Adjust Vertical Output Bias
a. Connect the High Frequency Constant-Amplitude
Signal Generator (067-0532-01) output to the CH 1 input
5-28 ®
via a GR-to-BNC female adapter and a 50 £2 BNC
termination.
b. Adjust the High-Frequency signal generator for a
200 MHz sine-wave output on the 0.5 V amplitude range.
c. ADJUSTVertical Output Bias adjustment, R488 (see
Fig. 5-5) for maximum deflection of the signal.
d. Disconnect the test setup.

Other manuals for Tektronix 475

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix 475 and is the answer not in the manual?

Tektronix 475 Specifications

General IconGeneral
BrandTektronix
Model475
CategoryTest Equipment
LanguageEnglish

Related product manuals