Commands Listed in Alphabetical Order
Arguments
ALWAYS continu
es looping until the self tests (diagnostics) are stopped via the
front panel or by an oscilloscope command.
FAIL causes lo
oping until the first self test (diagnostic) failure or until self tests
(diagnostics) are stopped.
ONFAIL caus
esloopingonaspecific test group as long as a FAIL status is
returned from the test.
ONCE execut
es self test (diagnostics test) sequence once.
NTIMES runs “n” number of loops.
Ex
amples
DIAG:LOOP:OPTION ONCE runs one loop of self tests.
DIAg:LOOP:OPTion:NTIMes
Sets the self-test loop option to run N times.
Group
Calibration a nd Diagnostic
Syntax
DIAg:LOOP:OPTion:NTIMes <NR1>
DIAg:LOOP:OPTion:NTIMes?
Arguments
<NR1> is the number of self-test loops.
Examples
DIAG:LOOP:OPTION:NTIMES 3 sets the self-test loop to run three times.
DIAG:LOOP:OPTION:NTIMES? might return :DIAG:LOOP:OPTION:NTIMES 5,
indicating that the self-test loop is set to run five times.
DIAg:LOOP:STOP (No Query Form)
Stops the self-test at the end of the current loop.
Group
Calibration a nd Diagnostic
Syntax
DIAg:LOOP:STOP
Examples
DIAG:LOOP:STOP stops the self test at the end of the current loop.
2-136 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual