EasyManuals Logo
Home>Tektronix>Test Equipment>MSO2024B

Tektronix MSO2024B Programmer's Manual

Tektronix MSO2024B
441 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #187 background imageLoading...
Page #187 background image
Commands Listed in Alphabetical Order
Group
Horizontal
Syntax
HORizontal:SCAle <NR3>
HORizontal:SCAle?
Arguments
<NR3> species the range from 2 ns to 100 s, depending on the oscilloscope model.
Examples
HORIZONTAL:SCALE 2E-6 sets the main scale to 2µs per division.
HORIZONTAL:SCALE? might return :HORIZONTAL :MAIN:SCALE 2.0000E-06
indicating that the main scale is currently set to 2 µs per division.
HORizontal:TRIGger:POSition
Sets the horizontal position when d elay mode is OFF. It is similar to
HORizontal:POSition.
Group
Horizontal
Syntax
HORizontal:TRIGger:POSitio n
HORizontal:TRIGger:POSitio n?
Examples
HORIZONTAL:TRIGGER:POSITIO N? might return
HORIZONTAL:TRIGGER:POSITIO N 50 indicating that the horizontal trigger
position is set to 50%.
ID? (Query Only)
Returns identifying information about the oscilloscope and related rmware.
Group
Miscellaneous
Syntax
ID?
Related Commands
*IDN?
Examples
ID? might return TEK/DPO2024,CF:91.1CT,FV:v1.0000. This indicates the
oscilloscope model number, congured format, and rmware version number.
MSO2000B, DPO2000B, MSO2000 and DPO2000 Series Oscilloscopes Programmer Manual 2-171

Other manuals for Tektronix MSO2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO2024B and is the answer not in the manual?

Tektronix MSO2024B Specifications

General IconGeneral
BrandTektronix
ModelMSO2024B
CategoryTest Equipment
LanguageEnglish

Related product manuals