Commands Listed in Alphabetical Order
MEASUrement:
MEAS<x>:MEAN? (Query Only)
Returns the mean value accumulated for this measurement since the last s tatistical
reset. Measurements are specied by x, which ranges from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MEAN?
Examples
MEASUREMENT:MEAS1:MEAN? might r eturn :MEASUREMENT:MEAS1: MEAN
514.71E
-09.
MEASUrement:MEAS<x>:MINImum? (Query Only)
Returns the minimum value for this measurement since the last statistical reset.
Measurements are specied by <x>, which ranges from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:MINIm um?
Examples
MEASUREMENT:MEAS1:MINIMUM ? might return
:MEASUREMENT:MEAS1:MINIMU M 1.75E-09.
MEASUrement:MEAS<x>:SOURCE[1]
Sets or returns the source for all single source measurements and species
the source to measure "from" when taking a delay measurement or phase
measurement. Measurements are specied by <x>, which ranges from 1 through 4.
Group
Measurement
Syntax
MEASUrement:MEAS<x>:SOURC E[1] {CH<x>|MATH<y>|REF<x>}
MEASUrement:MEAS<x>:SOURC E[1]?
Arguments
CH<x> is an input channel waveform, where x is the channel number.
MATH<y> is a math waveform, where y is 1.
REF<x> is a reference waveform, where x is the reference channel number.
2-200 MSO2000B, DPO2000B, M SO2000 and DPO2000 Series Oscilloscopes Programmer Manual

Other manuals for Tektronix MSO2024B

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO2024B and is the answer not in the manual?

Tektronix MSO2024B Specifications

General IconGeneral
BrandTektronix
ModelMSO2024B
CategoryTest Equipment
LanguageEnglish

Related product manuals