Performance Verication
3. Unzip the les a
nd copy the "rmware.img" le into the root folder of a
USB ash drive.
4. Power off your
oscilloscope.
5. Insert the USB ash drive into a USB Host port on the front or back of the
oscillosco
pe.
6. Power on the oscilloscope. The oscilloscope automatically recognizes the
replacemen
t rmware and installs it.
If the instrument does not install the rmware, rerun the procedure. If the
problem co
ntinues, contact qualied service personnel.
NOTE. Do not power off the oscilloscope or remove the USB ash drive until the
oscilloscope nishes installing the rmware.
The oscilloscope displays a message when the installation is complete.
7. Power o
ff the oscilloscope and remove the USB ash drive.
8. Power on the oscilloscope.
9. Push the Utility front-panel button.
10. Push the Utility Page lower-bezel button.
11. Turn multipurpose knob a and select Cong.
12. Push the About lower-bezel button. The oscilloscope displays the rmware
version number.
13. Conrm that the version number matches that of the new rmware.
18 MSO4000 and DPO4000 Series Specications and Performance Verication

Other manuals for Tektronix MSO4000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MSO4000 Series and is the answer not in the manual?

Tektronix MSO4000 Series Specifications

General IconGeneral
BrandTektronix
ModelMSO4000 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals