Cation Exchange Column
2155 2 Installation
IM/2155 Rev. G 3
2 Installation
2.1 Installation
The column is secured vertically by two mounting clips that are
bolted to a suitable panel or wall (Fig. 2.1).
Conductivity cells are used in conjunction with conductivity
transmitters mounted nearby. Different column/cell
configurations are used to suit the application (Fig. 2.2).
Fig. 2.1 Mounting Information
Dimensions in mm (in.)
385
(15.2)
Cation Column
complete with
quick release
couplings
Mounting Clip
Mounting Clip
Suitable for M5 screw
Suitable for M5 screw
Fig. 2.2 Typical Cation Column Arrangement
Change-over
Valve
Change-over Valve
1 Cell 1 Column
After’ cation
conductivity
measurement only
2 Cell 2 Column
Before’ and After’ cation
conductivity measurements with
in-service and standby columns
3 Cell 3 Column
Before’ and After’ cation
conductivity measurements
with in-service, backup and
rinse columns
Conductivity
Cell