Reportable and Reference Range
(
continued)
Calculated:
Reportable Reference Range
Test Units Range (arterial) (venous)
Hemoglobin/Hb g/dL 5.1 – 25.5 12 – 17 12 – 17
g/L 51 – 255 120 – 170 120 – 170
mmol/L 3.2 –15.8 7 – 11 7 – 11
TCO
2
mmol/L (mEq/L) 5-50 23 – 27 24 – 29
(for all cartridges but
the CHEM8+ only)
HCO
3
mmol/L (mEq/L) 1.0 – 85.0 22 – 26 23 – 28
BE mmol/L (mEq/L) (-30) – (+30) (-2) – (+3) (-2) – (+3)
Anion Gap/AnGap mmol/L (mEq/L) (-10) – (+99) 10 – 20 10 – 20
sO
2
% N/A 95 – 98
Test Flags and Operator Action
n ***: Results that are not reportable due to sensor errors or interfering substances.
Draw a fresh sample and repeat test. If results are flagged again, send sample to
the lab.
n < , > and < >: Results that are below or above the reportable range or dependant on
results that are outside the reportable range. Send sample to the lab if necessary.
n and : Results that are above or below the action range. Follow facility procedure
for samples with critical values.
9
Printing Test Results from a
i-STAT Printer or Martel Printer
Without Downloader or Downloader/Recharger
1. Turn printer on if green power light is not on.
2. Align IR windows of handheld and printer.
3. Display results.
4. Press .
5. Do not move handheld or printer until printing is complete.
6. If printer is not powered from a wall outlet, turn printer off.
With Downloader or Downloader/Recharger
1.
Place handheld in Downloader or Downloader/Recharger that is wired to printer.
2. Display results.
3. Press . Do not move handheld or printer until printing is complete.
4.
If printer is not powered from a wall unit using the AC adapter, turn printer off.
Printing more than one result
1. Turn the handheld on.
2. Press .
3. Press for Data Review.
4. Press for List.
5. Scroll through the test records using the and keys.
6. Press the numbered key for the test record(s). (Press the numbered key again to
deselect a record.)
7. Align handheld and printer IR window or place in Downloader or
Downloader/Recharger attached to printer. Press .
8. Do not move handheld or printer until printing is complete.
9.
If printer is not powered from a wall unit using the AC adapter, turn printer off
.
2
7
MENU

Other manuals for Abbott i-STAT 1

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Abbott i-STAT 1 and is the answer not in the manual?

Abbott i-STAT 1 Specifications

General IconGeneral
BrandAbbott
Modeli-STAT 1
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals