Checkout Information 2
Natural Gas Analyzers 16
8 meter CP-Sil 5 CB
The 8 meter CP-Sil 5 CB channel of the Natural Gas Analyzer A
Extended is used for the analysis of hydrocarbons from
n-hexane up to n-dodecane.
The NGA Gas Calibration standard and the Universal Gas
Calibration standard do not contain heavy hydrocarbons up to
n-dodecane. Reference chromatograms of the NGA Gas
Calibration standard and the Universal Gas Calibration
standard are included in this manual. For heavy hydrocarbons
up to n-dodecane an additional sample is analyzed for
identification.
Analysis of the NGA Gas Calibration standard results in the
chromatogram shown in Figure 12.
Reference check of the Universal Gas Calibration standard
results in a chromatogram similar to Figure 13.
Figure 12 NGA Gas Calibration standard analyzed with the 8 meter CP-Sil 5 CB channel on Analyzer A Extended
Seconds
15
20 25 30 35 40 45 50 55
mV
0
5
10
15
n-Hexane
Figure 13 Universal Gas Calibration standard analyzed with the 8 meter CP-Sil 5 CB channel on Analyzer
A Extended
Seconds
15
20 25 30 35 40 45 50 55
mV
0.0
2.5
5.0
7.5
n-Hexane
n-Heptane

Other manuals for Agilent Technologies 490 Micro GC

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 490 Micro GC and is the answer not in the manual?

Agilent Technologies 490 Micro GC Specifications

General IconGeneral
BrandAgilent Technologies
Model490 Micro GC
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals