EasyManuals Logo
Home>BLUM>Measuring Instruments>TC62

BLUM TC62 User Manual

BLUM TC62
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
Mounting and commissioning
by Blum-Novotest 2012 17
3.2 Mounting of the tool holder
CAUTION
Destruction of the measuring system
The probe is not allowed to be twisted for mounting of the tool holder.
Holding screw (A) must be aligned with the centering screws.
If the measuring system is in the working area we recommend visual contact to the
receiver.
1. Insert probe into tool holder (System BTH 50 resp. BTH 25)
2. Tighten screw (A) slightly only
3. Mount stylus (see chap. 3.4)
Adjustment of the run-out at TC61 with optical presetter!
4. Insert probe into spindle (TC60 / TC62 / TC63 / TC64 only)
5. Turn probe by 360°
6. Observe display of run-out stylus is not allowed to be deflected.
7. Adjust concentricity (4 internal hexagon screws 2 mm)
e.g. X-direction (fig. 3.4):
Release centering screws (1) and (3) slightly only.
Alignment by reciprocal release and tightening of centering screws (2) and (4).
Y-direction (fig. 3.5):
Release centering screws (2) and (4) slightly only.
Alignment by reciprocal release and tightening of centering screws (1) and (3).
Run-Out TC60 / TC62 / TC63 / TC64: < 10 µm
Run-Out TC61: < 100 µm
8. Tighten holding screw (A)
Concentricity adjustment is fixed now.
9. Check concentricity.
10. Carry out calibration cycle

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BLUM TC62 and is the answer not in the manual?

BLUM TC62 Specifications

General IconGeneral
BrandBLUM
ModelTC62
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals