Product specifications
Type: No touch & touch forehead thermometer (BNT 400)
Measuring range: 93.2 °F – 109.4 °F (34 °C – 43 °C)
Resolution: 0.1 °F (0.1 °C)
Laboratory Accuracy: 0.4 °F for 95 °F – 107.6 °F (0.2 °C for the range 35 °C – 42 °C)
outside that range ±0.5°F/0.3°C
(Ambient Temperature: 59 °F to 104 °F (15 °C to 40 °C)
This thermometer displays a calculated oral equivalent estimate.
Display: Liquid Crystal Display, 4 digits plus special icons
Acoustic: Audio:
Normal temp range = Green temp range: 1 long beep for 0.4 second duration
Fever = Red or Yellow temp range: 10 short beeps for 0.2 second duration
Operating temperature:
59
°
F - 104
°
F ( 15
°
C - 40
°
C )
Automatic Switch-o: Approx. 30 seconds after last measurement has been taken
Weight: 3.54 oz. (with batteries), 2.91 oz. (w/o batteries)
Long term storage ranges
Storage/transport temperature: -13 °F – 131 °F (-25 °C to 55 °C)
Humidity: 15–95% non-condensing
Battery: (2) AAA Batteries - at least 500 measurements
Pressure: 700-1060 hPA (0.7-1.06 atm)
This infrared thermometer meets requirements established in ASTM Standard E 1965-98 (for the
thermometer system). Full responsibility for the conformance of the product to the standard is
assumed by Kaz USA, Inc., a Helen of Troy Company, Health & Home, 400 Donald Lynch Blvd., Suite 300,
Marlborough, MA 01752
ASTM laboratory accuracy requirements for the thermometer only in the display range of 36 °C to 39°C
(96.8 °F to 102.2 °F) for infrared thermometers is ±0.3 °C (±0.5 °F), whereas for mercury-in-glass and
electronic thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ±0.1 °C (±0.2 °F).
This device conforms to the following standards:
EN 60601-1:2014 Medical electrical equipment. General requirements for basic safety and
essential performance
ASTM E1965-98:2016 - Standard Specication for Infrared Thermometers for Intermittent Determination
of Patient Temperature
ISO 80601-2-56:2017 Particular requirements for basic safety and essential performance of clinical
thermometers for body temperature measurement
IEC 60601-1-2:2014: Medical electrical equipment – Part 1-2: General requirements for basic safety and
essential performance – Collateral Standard: Electromagnetic disturbances – Requirement and tests
NOTE: Do not use this device in the presence of electromagnetic or other interference outside the normal
range specied in IEC 60601-1-2:2007.
EN 1041: 2008 Information supplied by the manufacturer of medical devices
IEC 60601-1-11: 2015 Medical electrical equipment – Part 1-11: General requirements for basic safety and
essential performance – Collateral standard: Requirements for medical electrical equipment and medical
electrical systems used in the home healthcare environment