10
Specifications
Model: BUA5000
Operation principle: Oscillometric method
Display: Liquid crystal display
Range of measurement: Pressure 30~280 mmHg
Pulse 40~199 beats/minute
Laboratory accuracy: ±3 mmHg (cuff pressure)
±5% of reading (pulse rate)
Clinical accuracy: <5 mmHg systematic offset
<8 mmHg standard deviation
Battery life: 400 measurements
Auto power off: Whenever not used for 1 minute
Accessories: 4 batteries, 1 arm cuff with tube,
instruction manual
Inflation: Automatic
Power supply: 4 batteries, type AA 1.5V
Operating temperature/ 10 °C to 40 °C (50 ˚F to 104 ˚F)
atmospheric pressure/ 860 to1060 hPa (0.849 to 1.046 atm)
humidity: 15% to 90% relative humidity, non-condensing
Storage temperature/ -20 °C to 55 °C (-4 ˚F to 131 ˚F)
atmospheric pressure/ 860 to1060 hPa (0.849 to 1.046 atm)
humidity: 15% to 90% relative humidity, non-condensing
Cuff: Universal cuff (22-42 cm)
Service life: 5 years
Warranty: 2 years
If device is not used within specified temperature and humidity ranges the
technical accuracy cannot be ensured.
Type BF applied parts
40
10
Operating temperature
See instructions for use
55
Storage temperature
Keep dry Storage humidity/
operating humidity
Subject to change without notice.
Internally powered equipment.
Continuous operation.
Protected against solid foreign particles with a diameter of more
than 12.5 mm, no protection against liquid.