How to set a default age group
You have the option to set a default AgeSmart™ age group. This will allow the device to power on
and be used in default selection without re-selecting with each power on.
1. When the device is o, hold both the power button and the AgeSmart™ button for 4 seconds,
the backlight will turn yellow, you can then release the buttons.
2. All Age Icons will ash continuously until the AgeSmart™ button is pressed.
3. Press the AgeSmart™ button, the 0-3 month icon will be filled in and ash and the other icons
will be empty. A second button press will advance to the 3-36 month icon ashing, a third will
advance to 36+months ashing. This cycle continues in order with each button press. A full
beep will sound with every button press.
4. After 3 seconds of no button press, the displayed icon will remain on the screen and a full beep
will sound. The device will turn o. When the device is turned back on, the default will be set.
Default will be set on the device until the batteries are removed and then put back in.
How to change temperature between °C and °F
Your Braun ThermoScan® is shipped with the Celsius (°C) temperature scale activated. If you wish to switch to
Fahrenheit (°F) and/or back from Fahrenheit to Celsius, proceed as follows:
1. Make sure the thermometer is turned o.
2. Press and hold down the Power button
. After about 3 seconds the display will show this
sequence: °C / SET / °F / SET.
3. Release the Power
button when the desired temperature scale is shown. There will be a
short beep to confirm the new setting, then the thermometer is turned o automatically.
Care and cleaning
The probe tip is the most delicate part of the thermometer. It must be clean and intact to ensure accurate
readings. If the thermometer is ever accidentally used without a lens lter, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. After the
alcohol has completely dried out, you can put a new lens filter on and take a temperature
measurement.
If the probe tip is damaged, please contact local service center.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not use abrasive cleaners.
Never submerge this thermometer in water or any other liquid. Store thermometer and lens filters
in a dry location free from dust and contamination and away from direct sunlight.
Additional lens filters (LF 40) are available at most stores carrying Braun ThermoScan®.
4 sec 3 sec
4 sec 3 sec
4 sec 3 sec
4 sec 3 sec
Replacing the batteries
1. Insert new batteries when the battery symbol appears on the display.
2. Open the battery compartment. Remove the batteries and replace with new batteries, making
sure the poles are in the right direction.
3. Snap battery door into place.
To protect the environment, dispose of empty batteries at appropriate collection site
according to national or local regulations.
Errors and troubleshooting
Error message Situation Solution
No lens filter is attached. Attach new, clean lens filter.
The thermometer probe was not positioned
securely in the ear.
An accurate measurement
was not possible.
ERROR + position error animated icon
Take care that the positioning of the probe is correct
and remains stable.
Replace lens filter and reposition.
Press Start button to begin a new measurement.
Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F).
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10 and 40 °C
or 50 and 104 °F .
Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2 – 108 °F).
HI = too high
LO = too low
Make sure the probe tip and lens are clean and a
new, clean lens filter is attached. Make sure the
thermometer is properly inserted. Then, take a new
temperature.
System error – self-check display ashes
continuously and will not be followed by
the ready beep and the ready symbol.
If error persists,
If error still persists,
Wait 1 minute until the thermometer turns o
automatically, then turn on again.
... reset the thermometer by removing the batteries
and putting them back in.
... please contact the Service Centre.
Battery is low, but thermometer will still
operate correctly.
Insert new batteries.
Battery is too low to take correct
temperature measurement.
Insert new batteries.
System error without messages or
completely dead batteries
Insert new batteries. If the error persists, call
customer service.
ThermoScan

Other manuals for Braun IRT6525

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Braun IRT6525 and is the answer not in the manual?

Braun IRT6525 Specifications

General IconGeneral
BrandBraun
ModelIRT6525
CategoryThermometer
LanguageEnglish

Related product manuals