Usage of Oxygen (Accessory)
Replacement of the Patient Air Inlet Filters
Symbols
Vivo 45
Quick Reference Guide
Doc. 007610 En Y-1
2019-10-29
•O
2
must be turned off when the Vivo 45 is not in
operating mode!
• It is recommended to monitor the oxygen concen-
trations.
• Oxygen flow must not exceed 15 l/min.
1. Mount the oxygen tube to
the connector.
2. Connect the oxygen tube. A
click sound is heard when the
tube is properly attached.
Disconnect:
1. Turn off the oxygen source.
2. Press the release button to
disconnect the oxygen tube.
3. Remove the tube.
The filters are located in the filter
cassette.
Washable Filter (grey): Replace at
least once a year, wash at least
once a week.
Disposable Filter (white): Replace
at least every 4th week, or more
often in high pollution or pollen-
rich environments.
SYMBOL DESCRIPTION
Internal battery
Click-in battery level
Home Mode activated
Ramp function active
Humidifier and/or heated circuit connected
SpO2 sensor connected
FiO2 connected
EtCO2 connected
PtcCO2 connected
Effort belt connected