EasyManuals Logo
Home>Dräger>Test Equipment>Alcotest 7000

Dräger Alcotest 7000 Manual

Dräger Alcotest 7000
10 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #9 background imageLoading...
Page #9 background image
Technical Data
Dräger Alcotest
®
7000 | 09
Dräger Alcotest 7000
Detection basis
Measurement range
Electrochemical DrägerSensor
®
, 1/4" technology, alcohol-specific
Breath alcohol concentration: 0 to 3.0 mg/l;
(Ethanol mass per minimum volume at 34 °C and 1,013 hPa)
Blood alcohol concentration: 0.00 to 6.00 (Ethanol mass per
blood volume or mass at 20 °C and 1013 hPa)
Visual signal when measuring limits are exceeded
Sample collection Default: Automatic sample collection when minimum volume or
predefined blowing time is reached;
Passive sampling or manual triggering of sampling possible
Start-up time Approx. 3 seconds after switching on (depending on
configuration)
Display of test results After approx. 3 seconds (at 0.0 mg/l)
After approx. 10 seconds (at 0.5 mg/l, at room temperature)
Ambient operating conditions -10 °C to +55 °C; 10 to 100 % relative humidity
(non condensing); 600 to 1,300 hPa
Ambient conditions during storage -20 °C to +60 °C; 15 to 75 % relative humidity
(10 to 100% relative humidity <48 h); 600 to 1,300 hPa
Display QVGA TFT colour display 320 x 240 2.4"
LEDs tricolour LEDs to help display results and warning messages
Acoustic signals Various signal tones to support display and warning messages
Memory Storage of the last 30,000 tests (minimum information) or
10,000 tests (high-detail information)
Energy supply Lithium-ion battery (rechargeable via USB-C connection) or
alkaline version (3 x 1.5 V alkaline type or 3x NiMH or 3x lithium
primary cell (Mignon, LR06, AA),
Charge status shown in display; one charge is sufficient to
perform approx. 5,000 breath test (lithium version) or
1,500 breath tests (alkaline version).
Mouthpiece receptacle The Slide ’n’ Click mouthpiece or Alcotest funnel mounts on either
side
Mouthpiece Hygienic, individually packaged with tamper-proof, unclosable air
outlet
Mouthpiece ejector and spacer between mouth and device
housing
Funnel Hygienic, individually packaged with air outlets for minimising
back-flow of expired air
Operating design Measuring functions can be performed with one key;
Navigation in menu with two menu keys
Calibration Wet gas or dry gas calibration;
Recommended calibration interval of 12 months
Casing Impact-resistant ASA/PC
Dimensions (W x H x D), Weight 61 mm x 172 mm x 36 mm;
A7000 LI: 275 g
A7000 AA: 285 g
Device configuration Configuration of device settings direct via device menu
(PIN required)
No additional PC software required
Vibration and impact resistance EN 60068-2-27, EN 60068-2-6; EN 60068-2-64
CE mark Electromagnetic compatibility
Standards EN 15964, NHTSA, FDA-conformity, depending on configuration
Internal clock Warning or switch-off after expiry of calibration period

Other manuals for Dräger Alcotest 7000