General Specifications
Clamp jaw opening 52mm (2.0") approx.
Display Dual 40,000/4,000 count backlit LCD
Continuity check Threshold 25 to 60; Test current < 0.5mA
Diode test Test current of 0.3mA typical;
Open circuit voltage 2.8VDC typical
Low Battery indication Battery symbol is displayed
Over-range indication ‘OL’ display
Measurement rate 2 readings per second, nominal
Peak detector >1ms
Thermocouple sensor Type K thermocouple required
Input Impedance 10M (VDC and VAC)
AC bandwidth 50 to 400Hz (AAC and VAC)
AC response True rms (AAC and VAC)
Crest Factor 3.0 in 400A range, 1.4 in 1000A range (50/60Hz and 5% to 100% of
range)
Operating Temperature 5C to 40C (41F to 104F)
Storage Temperature -20C to 60C (-4F to 140F)
Operating Humidity Max 80% up to 31C (87F) decreasing linearly to 50% at 40C
(104F)
Storage Humidity <80%
Operating Altitude 2000meters (7000ft.) maximum.
Battery One (1) 9V Battery (NEDA 1604)
Auto power OFF After approx. 30 minutes
Dimensions & Weight 294x105x47mm (11.57x4.13x1.85”); 536g (18.9 oz)
Safety For indoor use and in accordance with the requirements for double
insulation to IEC1010-1 (2001): EN61010-1 (2001) Overvoltage
Category IV 600V and Category III 1000V, Pollution Degree 2.
Approvals CE
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Other manuals for Extech Instruments MA1500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Extech Instruments MA1500 and is the answer not in the manual?

Extech Instruments MA1500 Specifications

General IconGeneral
BrandExtech Instruments
ModelMA1500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals