Stik Thermometer
SCPI Commands
13
Battery Life
Battery life is approximately 300 hours (12.5 days) of continuous operation with the
backlight off. A low-battery icon (E) shows in the lower right of the display when
battery level is low. Replace batteries per recommendations found in the specifications
section of this manual.
Calibration
For calibration information, refer to the document on the documentation CD.
SCPI Commands
For a list of the SCPI commands, refer to the document on the documentation CD.
RS-232 Interface
An RS-232 interface is standard on the Product. Serial communication can be used for
configuration, calibration, and to move measurement data from the Product. An RS-232
cable is included with data logging software purchase.
XWWarning
To prevent possible electrical shock, fire, or personal injury, the
RS-232 interface must not be used in hazardous areas.
Specifications
(Ambient: 23 °C ±5 °C)
Measurement Range
1551A Ex ......................................................................... 50 °C to 160 °C (58 °F to 320 °F)
1552A Ex ......................................................................... 80 °C to 300 °C (112 °F to 572 °F)
Accuracy (1 year) ..................................................................... ±0.05 °C (0.09 °F)
Resolution ................................................................................. Selectable (0.1, 0.01, 0.001) factory default is 0.01
Sample Rate ............................................................................. User selectable 0.5/sec, 1/sec or 2/sec factory default is
1/sec
Operating Temperature Range of Readout: .......................... 10 °C to 50 °C (14 °F to 122 °F)
Probe Response Time .............................................................. Approximately 20 seconds
EMC Compliance....................................................................... EN61326:2006 Annex C
CISPR 11, Edition 5.0-2009
Class “B”
Humidity Range......................................................................... 0 to 95 % RH Non-condensing
Storage Temperature Range.................................................... 20 °C to 60 °C (4 °F to 140 °F)
Enclosure Rating ...................................................................... IP50
Readout Temperature Coefficient ........................................... Add ±10 ppm/°C of full scale temperature from 10 °C to
18 °C and 28 °C to 50 °C
Probe Temperature Coefficient .............................................. 0.00385 Ω/Ω/°C nominal
Nominal Probe Resistance at 0 °C .......................................... 100 Ω
Probe Hysteresis....................................................................... ±0.01 °C
Power ......................................................................................... 3 AAA alkaline batteries
(Must use only approved batteries. See Table 2)

Other manuals for Fluke 1551A Ex

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fluke 1551A Ex and is the answer not in the manual?

Fluke 1551A Ex Specifications

1551A Ex

Measurement Specifications IconMeasurement Specifications
measurement range−50 °C to 160 °C
accuracy±0.05 °C
resolutionSelectable (0.1, 0.01, 0.001) factory default is 0.01
Environmental Specifications IconEnvironmental Specifications
operating temperature range−10 °C to 50 °C
humidity range0 to 95 % RH Non-condensing
storage temperature range−20 °C to 60 °C
Power Specifications IconPower Specifications
power source3 AAA alkaline batteries
battery lifeApproximately 300 hours without back light
battery save rangeSelectable from 1 to 30 minutes or can be disabled
Weight and Dimensions IconWeight and Dimensions
weight200 g
size (readout only)125 mm x 62 mm x 35 mm
probe size 1551A-94.8 mm x 229 mm

1552A Ex

Measurement Specifications IconMeasurement Specifications
measurement range−80 °C to 300 °C
accuracy±0.05 °C
resolutionSelectable (0.1, 0.01, 0.001) factory default is 0.01
Environmental Specifications IconEnvironmental Specifications
operating temperature range−10 °C to 50 °C
humidity range0 to 95 % RH Non-condensing
storage temperature range−20 °C to 60 °C
Power Specifications IconPower Specifications
power source3 AAA alkaline batteries
battery lifeApproximately 300 hours without back light
battery save rangeSelectable from 1 to 30 minutes or can be disabled
Weight and Dimensions IconWeight and Dimensions
weight200 g
size (readout only)125 mm x 62 mm x 35 mm
probe size 1552A6.35 mm x 305 mm