16
Power Failure
What happens when power is restored?
• Yourcyclerwillgothrougha“PowerFailRecovery”processifpowerfailhappenedduring
drain, fill, dwell or pause
• Anaudiblebeepwillbeheard.PressSTOP to mute the warning.
• Yourcyclerwillperformasafetycheck.Thestatusbarwillstate“PowerFailRecovery”
and the screen will read “Power Fail Detected Recovery Check in progress”.
• Ifyourcyclerpassesthesafetycheckitwillbeepagain.
Press STOP to mute and then you may select “Resume Treatment”.
• Ifyourcyclerdoesnotpassthesafetycheck,select“CancelTreatment”
and follow the prompts on the screen to start your treatment over.
Use new supplies for new treatment.
NOTE: If the power has been off for more than
8 hours do NOT resume treatment. Discuss with
your PD Nurse any special instructions in case of
a power failure.