1. Make sure that the fiberoptic cable
connector is fully inserted into the light box.
2. Make sure that the surface of the fiberoptic
light pad is not damaged or discolored.
3. Make sure that the fiberoptic lens has not
been damaged or discolored.
4. Make sure that the average light output
measurement is taken with the light pad on a
flat surface and that the pad is not covered
with a BiliSoft Pad Cover or Nest.
5. Make sure to use a calibrated Ohmeda
Medical BiliBlanket Meter II to take the light
output readings.
6. Perform the comprehensive 6-point (small
pad) or 9-point (large pad) check.
7. If the light output is still 25% below minimum
rated value, remove the unit from service
and seek assistance from service personnel.
1. Remove the unit from service and seek
assistance from service personnel.
2. All damaged parts must be replaced and unit
checked to ensure no internal damage has
occurred.
1. The hour meter only runs when the fiberoptic
light pad is illuminated. If the fiberoptic cable
connector is not fully inserted in the light box,
the LED module is automatically shut off and
the hour meter does not run.
2. If the fiberoptic cable connector is fully
inserted and the hour meter still does not
work, remove the unit from service and seek
assistance from service personnel.