EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ E10

GE LOGIQ E10 Technical Publication

GE LOGIQ E10
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #55 background imageLoading...
Page #55 background image
General Ultrasound system requirements
Ultrasound SystemCommon Service Information 3-3
Direction 5444964-100 English Rev. 5
General Ultrasound system
requirements
Contents in this section
• ‘Ultrasound system environmental requirements’ on
page 3-3
• ‘Electrical requirements’ on page 3-5
• ‘EMI limitations’ on page 3-18
‘EMI prevention/abatement’ on page 3-19
‘Probes environmental requirements’ on page 3-20
‘Time and manpower requirements’ on page 3-20
Ultrasound system environmental requirements
If the Ultrasound system is very cold or hot
When unpacking the Ultrasound system, allow the temperature
of the Ultrasound system to stabilize before powering up. The
following table describes guidelines for reaching operational
temperatures from storage or transport temperatures.
CAUTION
If the Ultrasound system is very cold or hot, do not turn on its
power until it has had a chance to acclimate to its operating
environment.
Table 3-1: Ultrasound system acclimate time

Table of Contents

Other manuals for GE LOGIQ E10

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ E10 and is the answer not in the manual?

GE LOGIQ E10 Specifications

General IconGeneral
BrandGE
ModelLOGIQ E10
CategoryMedical Equipment
LanguageEnglish

Related product manuals