Electrical safety tests
Ultrasound SystemCommon Service Information 4-41
Direction 5444964-100 English Rev. 5
Isolated Probe (Sink) Leakage-Isolation Test (continued)
Record all data and keep the record of the result with other hard
copies of maintenance data.
NOTE: Not all test procedures are applicable to all areas of the world.
Reversed Polarity testing content satisfies regions following IEC
62353:2007 and IEC 60601-1:2005.
NOTE: Probe names in italics font are given as examples only.
Table 4-18: Typical data format for recording Isolated Probe (sink) leakage
Unit under test___________________________ Date of test:___________
Test Conditions Probe as measured in saline bath
System Power Grounding /PE 4C i12L TS E8C
off closed
on closed
Table 4-19: Typical data format for recording Isolated Probe (sink) leakage
Unit under test____________________________________ Date of test:_____________
Test Conditions Probe as measured in saline bath
System
Power
Grounding/
PE
Limit
µA 4C i12L TS E8C
off closed 5000
on closed 5000
off closed
(reversed
polarity)
5000
on closed
(reversed
polarity)
5000

Table of Contents

Other manuals for GE LOGIQ P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P6 and is the answer not in the manual?

GE LOGIQ P6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals