EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ P6

GE LOGIQ P6 Technical Publication

GE LOGIQ P6
127 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #80 background imageLoading...
Page #80 background image
Care and Maintenance
4-2 Ultrasound SystemCommon Service Information
Direction 5444964-100 English
Rev. 5
Overview
Care and Maintenance
Normal Care and Maintenance is mandatory; it is the
responsibility of the customer and includes the following:
Quality Assurance Program to perform routine quality
control testing.
NOTE: Some customers’ Quality Assurance Programs may require
additional tasks and or inspections at a different frequency
than listed in this manual.
System Maintenance and Checks (Preliminary System,
Functional Checks, Peripheral, Cable, Physical Inspection).
• Probe Maintenance
System Cleaning, including Air Filter Cleaning
Electrical Safety Testing
NOTE: It has been determined by engineering that your Ultrasound
System DOES NOT have any high wear components that fail
with use. Therefore, Periodic Maintenance inspections are
not mandatory for any Ultrasound system. Only normal Care
and Maintenance is recommended.
Contents in this chapter
• ‘Warnings’ on page 4-3
‘Why do maintenance’ on page 4-4
‘Maintenance task schedule’ on page 4-5
‘System maintenance’ on page 4-6
‘Using a Phantom’ on page 4-13
‘Electrical safety tests’ on page 4-14
‘When there's too much leakage current …’ on page 4-44
• ‘Ultrasound Equipment Quality Check (EQC and IQC)’ on
page 4-46

Table of Contents

Other manuals for GE LOGIQ P6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ P6 and is the answer not in the manual?

GE LOGIQ P6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ P6
CategoryMedical Equipment
LanguageEnglish

Related product manuals