Electrical safety tests
Ultrasound SystemCommon Service Information 4-25
Direction 5444964-100 English Rev. 5
Chassis leakage current test
Definition
This test, also known as Enclosure Leakage current test,
measures the current that would flow through a grounded
person who touches the accessible conductive parts of the
equipment during normal and fault conditions.
The test verifies the isolation of the power line from the chassis.
The testing Electrical Safety Analyzer is connected to parts of
the equipment, easily contacted by the user or patient.
Record the highest reading.
DANGER Electric Shock Hazard.
When the Electrical Safety Analyzer's ground switch is OPEN, DO NOT
touch the Ultrasound system!
CAUTION
Equipment damage possibility.
Never switch the Polarity and the status of Neutral when the
Ultrasound system is powered ON.
Power off the Ultrasound system, allow the stored energy to
bleed down, and turn the circuit breaker off BEFORE switching
the "POLARITY" switch and/or the "NEUTRAL" switch on the
Electrical Safety Analyzer to avoid possible power supply
damage.

Table of Contents

Other manuals for GE LOGIQ S6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ S6 and is the answer not in the manual?

GE LOGIQ S6 Specifications

General IconGeneral
BrandGE
ModelLOGIQ S6
CategoryMedical Equipment
LanguageEnglish

Related product manuals