General Procedures and Functional Checks
4-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
4-2-7 Backup and Restore Database, Preset Configurations and
Images
NOTE: Always save presets before any software reload. This ensures
the presets loaded after the software reload are as up–to–date
as possible.
All user presets except changes to Summary, Anatomy, and
Biometry pages, can be saved on an DVD-R disk (or USB
memory device) for reloading on the system.
NOTE: Presets should NOT be saved on the same USB memory device
(or DVD-R disk) as images. The Archive Menu lists the images
but does NOT list the presets stored on a USB memory device
(or DVD-R disk).