General Procedures and Functional Checks
4-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
4-3-9 2D Mode (B mode) Checks
4-3-9-1 Introduction
The 2D Mode is the system’s default mode.
4-3-10 2D Mode Screen Example
4-3-11 Preparations
• Connect one of the probes.
• Turn ON the scanner.
The 2D Mode is displayed (default mode).
4-3-12 Adjust the 2D mode controls
Press B Mode on the Operator Panel to access 2D mode.
These Image Controls are used to optimize the 2D picture.
Verify that all the listed controls are working as intended:
• Use Gain and TGC controls to optimize the overall image
together with the Power control.
• Use Depth to adjust the range to be imaged.
• Use Focus to center the focal point(s) around the region of
interest.
• Use Frequency (move to higher frequencies) or Line
Density (move to higher line density) to increase resolution
in image.
• Use Frequency (move to lower frequency) to increase
penetration.
• Use the Rejection control to reduce noise in the image.

Table of Contents

Other manuals for GE LOGIQ V5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V5 and is the answer not in the manual?

GE LOGIQ V5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V5
CategoryMedical Equipment
LanguageEnglish

Related product manuals