General Procedures and Functional Checks
4-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
4-2-7-4 Archiving Images
1. Insert the archive media.
2. To format the archive media, enter Utility-> Connectivity->
Removable Media.
3. Format the CD. Verify the format if desired.
4. Images will be moved from the hard drive by date.
Therefore, the best way is to label media by date.
NOTE: Images will be moved from the hard drive by date.
Therefore, the best way to label media is by date. When
images are moved to the archive media, they will be deleted
from the system hard drive. However, the patient database
(backed up earlier) maintains pointers to the location of the
images on the archive media.
Figure 4-9. Format Media Screen
5. Enter Utility-> System-> Backup/Restore.
6. Select “Move File Older Than in Days”.