EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V5

GE LOGIQ V5 Service Manual

GE LOGIQ V5
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #41 background imageLoading...
Page #41 background image
Safety considerations
LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-17
5726264-100 English Rev.8
1-6-3 Human Safety(continued)
WARNING
Because of the limited access to cabinets and equipment in the
field, placing people in awkward positions, GE has limited the
lifting weight for one person in the field to 16 KG (35 LBS).
Anything over 16 KG (35 LBS) requires 2 people.
WARNING
Have two people available to deliver and unpack the LOGIQ
V3/V5/V5 Expert.
Attempts to move the Ultrasound system considerable
distances or on an incline by one person could result in injury
or damage or both.
WARNING
Ensure that the Ultrasound system is turned off and unplugged
Wait for at least 20 seconds for capacitors to discharge as
there are no test points to verify isolation. The amber light on
the OP panel ON/OFF button will turn off.
Ultrasound system components may be energized. Always
refer to the Ultrasound system's Proprietary Service Manual for
LOTO warnings and cautions

Table of Contents

Other manuals for GE LOGIQ V5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V5 and is the answer not in the manual?

GE LOGIQ V5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V5
CategoryMedical Equipment
LanguageEnglish

Related product manuals