EasyManuals Logo
Home>GE>Medical Equipment>LOGIQ V5

GE LOGIQ V5 Service Manual

GE LOGIQ V5
305 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #64 background imageLoading...
Page #64 background image
Site Preparations
2-8 LOGIQ V3/V5/V5 Expert – Basic Service Manual
5726264-100 English Rev.8
2-2-3-7 Power stability requirement
Table 2-4: Power stability requirement
IEC 61000-4-11
Voltage dips, short
interruptions and
voltage variations
on mains supply
< 5%T (> 95% dip) for
0.5 cycle;
< 5%T (> 95% dip) for
0.5 cycle;
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
40%T (60% dip) for 5
cycles;
40%T (60% dip) for 5
cycles;
70%T (30 dip) for 25
cycles;
70%T (30 dip) for 25
cycles;
< 5%T (>95% dip) for 5
sec
< 5%T (>95% dip) for 5
sec

Table of Contents

Other manuals for GE LOGIQ V5

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE LOGIQ V5 and is the answer not in the manual?

GE LOGIQ V5 Specifications

General IconGeneral
BrandGE
ModelLOGIQ V5
CategoryMedical Equipment
LanguageEnglish

Related product manuals